Design and Implement of Signal Generator in Magnetic Focused Conductivity Tomography System. [J]. Advanced Engineering Sciences 38(2):155-158(2006)DOI:
Design and Implement of Signal Generator in Magnetic Focused Conductivity Tomography System
Abstract:To improve stability of the signal source and perform sweeping detection
as well as extract abundance and reliable signal
the direct digital synthesis technology was employed to design the generator of source which formes a sweeping frequencies sine wave output from 1 MHz to 20 MHz. The planar spiral coil was connected as an amplitude modulation circuit. The same coil was made use of differential architecture for signal detection and extraction. The MC1595 was utilized to compose a phase detector
in which difference of phases varies with change of frequencies. A low pass filter was designed to filter the carry waves of the sweeping source. Thereby the system obtained abundance data and its stability was improved. Further
the spatial resolution of the system was enhanced. All of the above is in favor of using software in magnetic focused conductivity tomography system to reconstruct the image of conductivity within human body.
关键词
磁聚焦电导率成像信号源平面螺旋聚焦线圈乘法器信号提取
Keywords
magnetic focused conductivity tomography (MFCT)signal resourcefocused coilmultipliersignal extraction
references
Cole K S,Cole R H.Dispersion and absorption in dielectrics[J].J Chem Phys,1941,9(4):341-351.
Cole K S.Electric impedance of suspensions of spheres[J].J Gen Physiol,1928,12(1):29-36.
Lynn W H,Harvey W K,James H M,et al.A Noninvasive electromagnetic conductivity sensor for biomedical application[J].IEEE Transactions on Biomedical Engineering,1988,12(35):1011-1021.
Guo R X,Guo J X.Principle and application of 125 MHz DDS frequency synthesizer AD9850[J].Application of electronic technique,1997,23(12):65-67.[郭荣祥,郭吉祥.AD9850 125MHz DDS 频率合成器的原理及应用[J].电子技术应用,1997,23(12):65-67.]