罗教明,郑立,包崇云.多段式骨植入体周围骨应力分布的三维有限元分析[J].工程科学与技术,2006,38(4):53-57.
3D FEA of Bone Stress Around Multi-part Osseointegrated Implant. [J]. Advanced Engineering Sciences 38(4):53-57(2006)
中文摘要: 针对植入性下肢假肢采用传统的一段式骨植入体,长期植入下由于对植入体周围的骨产生应力屏蔽效应继而引起骨质疏松,从而导致植入体松动、脱落而失效的问题
设计了一种新型的多段式结构的植入体。基于人体CT扫描数据建立了相关股骨和骨内植入体模型,针对低、中和高位下肢截肢步行最大生理载荷情况下,采用三维有限元分析方法,对新设计周围骨应力进行了分析研究,并与传统一段式骨植入体和健康自然骨的应力分布情况进行了比较。结果表明新型植入体植入后相比于一段式应力屏蔽和集中的程度得到了有效减轻,与自然状态下骨的应力较为接近,低、中位截肢下应用较好,高位下应力集中较明显,并能保持其在体内的稳定性。
Abstract:Faced with the problem of the loose and extrusion
and then the malfunction of the osseointegrated implant for percutaneous osseointegrated artificial lower limb attachment after long-time of use caused by stress shielding
which occurred always around traditional one part implant
a new type of implant—multi part implant was developed. Based on CT data and under the maximal load during a normal walking cycle
3D finite element analysis was carried out to analyze the stress of bone around the new implant in three cases of distally truncated femur at high-position、middle-position and low-position. Results showed that stress shielding and stress concentration under the new type of implant are reduced effectively compared with the traditional one-part implant
and the stress distribution is much close to the natural bone. Application for distally truncated femur at middle-position and low-position was much better
while stress concentration was marked at high-position. Meanwhile
the stability in vivo can also be maintained with the multi part implant.The new implant can markedly reduce the adverse effect of stress shielding
which is promising applied for prosthetic limb.
multi-part osseointegrated implantstress shielding3D finite element analysispercutaneous osseointegrated artificial limb
multi-part osseointegrated implantstress shielding3D finite element analysispercutaneous osseointegrated artificial limb
Xu W,Crocombe A D,Hughes S C.Finite element analysis of bone stress and strain around a distal osseointegrated implant for prosthetic limb attachment[J].Proc Inst Mech Eng,2000,214(6):595-602.
Marks L J,Michael J W.Clinical review:science,medicine and the Future artificial limbs[J].BMJ,2001,323:732-735.
Liu JianGuo.The definition and research methods of the effect of stress shieldin[J].Journal of Medical Biomechanics,1992,7(2):136.[刘建国.应力遮挡效应的概念及研究方法[J].医用生物力学,1992; 7(2):136.]
Harris L J,Tarr R R.Implant failures in orthopaedic surgery[J].Biomater Med Device Artif Organs,1979,7(2):243-255.
Harrie Weinans,Dale R Sumner.Finite elementanalyses to study periprosthetic bone adaption[J].Bone Research in Biomechanics,1997,40:3-16.
Sloten J V,Hobatho M C,Verdonck P,et al.Application of computer modeling for the design of orthopaedic,dental and cardiovascular biomaterials[J].Proc Instn Mech Engres,1998,212:489-500.
Gross S,Abel E W.A finite element analysis of hollow stemmed hip prostheses as a means of reducing stress shielding of the femur[J].J Biomech,2001,34(8):995-1003.
Chinglung Tai,Chunhsiung Shih,Wengpin Chen,et al.Finite element analysis of the cervico-trochanteric stemless femoral prosthesis[J].Clin Biomech,2003,18(6):53-58.